Accuracy Specifications Function Range Resolution Accuracy AC Volts1 6.000 V

60.00 V
600.0 V 0.001 V
0.01 V
0.1 V 1.0 % + 3 DC Volts 6.000 V
60.00 V
600.0 V 0.001 V
0.01 V
0.1 V 0.5 % + 3 AC Millivolts 600.0 mV 0.1 mV 3.0 % + 3 Diode Test2 2.000 V 0.001 V 10 % Resistance 
(Ohms)
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ 0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ 0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5% + 3 Capacitance3 50.00 nF
500.0 nF
5.000 μF
50.00 μF
500.0 μF
1000 μF 0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
1 μF 2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
5 % + 5 Frequency4 Hz
(10 Hz – 100 kHz)
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz 0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz NA Duty Cycle4 1 % to 99 % 0.1 % NA AC Current
(40 Hz to 200 Hz)
4.000 A
10.00 A 0.001 A
0.01 A 1.5 % + 3 DC Current 4.000 A
10.00 A 0.001 A
0.01 A 1.5 % + 3

  1. All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
  2. Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA.
  3. Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 50 nF range).
  4. All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.

Function Overload Protection Input Impedance (Nominal) Common Mode Rejection Ratio Normal Mode Rejection Ratio AC Volts 600 V 1 >10 MΩ <100 pF 2 >60 dB at dc,
50 Hz or 60 Hz − AC Millivolts 600 mV >1M, <100 pF >80 dB at 50 Hz or 60 Hz − DC Volts 600 V 1 >10 MΩ <100 pF >100 dB at dc,
50 Hz or 60 Hz >60 dB at 50 Hz or 60 Hz

  1. 6 x 105 V Hz Max
  2. For mV (AC), input impedance is approximately 1 MΩ.

General Specifications Maximum Voltage Between any Terminal and Earth Ground 600 V Display (LCD) 6000 counts, updates 3/sec Battery Type 2 AAA, NEDA 24A, IEC LR03 Battery Life 200 hours minimum Temperature Operating 0 °C to 40 °C Storage -30 °C to 60 °C Relative Humidity Operating Humidity Non-condensing when <10°C
≤90 % at 10 °C to 30 °C; ≤75 % at 30 °C to 40 °C Operating Humidity, 40 MΩ Range ≤80 % at 10 °C to 30 °C; ≤70 % at 30 °C to 40 °C Altitude Operating 2000 m Storage 12,000 m Temperature Coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C) Fuse Protection for Current Inputs 11A, 1000V Fast Fuse, Fluke specified part only

 

Size (HxWxL) 142 mm x 69 mm x 28mm Weight 200 g IP Rating IEC 60529: IP 40 Safety IEC 61010-1: 600 V CAT III, Pollution Degree 2 Electromagnetic Environment IEC 61326-1: Portable Electromagnetic Compatibility Applies to use the Korea only.  Class A Equipment (Industrial Broadcasting and Communication Equipment)1